CleverPoint

Evidence / Device

Device validation

First the bench, then people.

Before a measure can mean anything, the device has to record the signal faithfully. We tested CleverPoint 4 in two steps: on a signal generator, where the input is known exactly, and on people, with the classic physiological checks for each signal it records.

Internal test protocol, version 3.2, November 2022: CleverPoint 4 board revision 4.5, EVA 2.102.

1. On the bench

A signal generator fed sine waves of known frequency and amplitude, from 1 to 100 µV and from 4 to 46.5 Hz, into the electrode inputs. Every channel was then compared with what went in.

Generator control window: eight channels of CleverPoint 4 showing sine waves of about 1 µV at 4, 7, 11, 15, 18, 26 and 46.5 Hz
Each channel receives its own frequency at about 1 µV.
The same eight channels as spectra: one clean peak per channel at the generator's frequency
The spectrum of each channel: one clean peak, where the generator put it.

Frequency response

Test signals from 4 to 46.5 Hz, covering the theta, alpha and beta rhythms and the frequencies that carry the timing of each heartbeat, come through as single clean peaks over a noise floor that stays flat up to about 47 Hz, both at 100 µV and at 1 µV, the smallest input tested.

Amplitude-frequency response at 100 µV: a single peak at the test frequency over a flat floor up to 47 Hz
Amplitude-frequency response at 1 µV: the test peak still stands clear of a flat floor

Sensitivity

How much of the input amplitude the module returns rises with the input: 25 % at 1 µV, then 80 % at 10 µV, 89 % at 20 µV, 95 % at 50 µV, 98 % at 100 µV. Scalp EEG lies mostly above 10 µV. Heart signals are measured in hundreds of microvolts and more.

Sensitivity curve: share of the input signal returned, rising from 25 % at 1 µV to 80 % at 10 µV and 98 % at 100 µV

2. On people

On the bench the input is known; on a person it is not. So each signal was checked against an effect every physiologist knows to look for.

EEG: the brain follows a flicker

For this check the electrodes were taken off the mask and placed at O1 and O2, over the visual cortex, the standard positions for this paradigm. During light flicker from 4 to 20 Hz the recording picks up the stimulation frequency: at 10 Hz a regular 10 Hz rhythm builds up, and the spectrogram shows a band at the flicker frequency for as long as it lasts. This is photic driving, the classic check that an EEG channel records brain rhythms.

The channel names in the screenshots (F7, AF3, AF4, F8) are the names of the module's inputs; in this recording the inputs were wired to electrodes at O1 and O2.

CleverPoint View: four module inputs wired to electrodes at O1 and O2 during 10 Hz photic stimulation, a regular 10 Hz rhythm building up on all four
10 Hz flicker, electrodes at O1 and O2: the rhythm builds up on every input.
Spectrograms of two module inputs at O1 and O2 during photic stimulation from 4 to 20 Hz: bands at the flicker frequency, stimulation periods marked below
Spectrograms of two inputs at O1 and O2: a band follows the flicker frequency.

ECG: the full PQRST shape

The ECG electrode on the forearm or the back of the hand records the complete cardiac cycle: P wave, QRS complex and T wave, with the R peaks sharp enough for beat-to-beat timing.

ECG recorded by CleverPoint 4: three cardiac cycles with P, QRS and T waves, R peaks marked

Blinks

Blinks show on the frontal channels as the characteristic large, smooth deflections. Stressonika counts them in each task and shows them as an observation.

Frontal channel: three blinks as large smooth deflections of about 50–75 µV

EMG: the forehead works with the task

The first nine minutes of a session on a frontal channel, with the sections marked above. Muscle activity is higher in the boxed stretches, which include tapping and both Stroop tasks, and drops in the rest break that follows.

The first nine minutes of a session on a frontal channel: higher muscle activity in the boxed stretches, including tapping and the Stroop tasks, lower in the rest break that follows

3. Beat detection

CleverPoint 4 reports each heartbeat interval itself, and our analysis also finds the R peaks in the raw ECG separately, with its own detector. On 30 sessions the two agree to a median of 0.26 bpm in heart rate (90 % of comparisons within 1.26 bpm).

4. Clean, and show the cleaning

Before any measure is computed, the beat series is checked for missed and extra beats, which are corrected. In a typical section nothing needs correcting (median 0 % of beats; mean 2.8 %; 95 % of sections need under 15 %). The report shows the share of corrected beats and adds a quality note when a recording needs one, so the reader always knows what the numbers rest on.

Next: how well the measures repeat →

Look at the raw signal yourself.